  "id": "leeds-phantom-unet-v1",
  "name": "Leeds phantom U-Net",
  "version": "1.0.0",
  "task": "Phantom anatomy and artifact segmentation",
  "clinically_validated": false,
  "patient_outcome_prediction": false,
  "source_domain": "lung ultrasound phantom",
  "license": "CC-BY-4.0",
  "source": "https://archive.researchdata.leeds.ac.uk/1263/",
  "attribution": "McLaughlan et al., University of Leeds, 2024. Dataset and derived model: CC BY 4.0.",
  "class_names": [
    "Background",
    "Ribs",
    "Pleural line",
    "A-line",
    "B-line",
    "B-line confluence"
  ],
  "colors": [
    "#000000",
    "#f59e0b",
    "#22d3ee",
    "#a78bfa",
    "#34d399",
    "#fb7185"
  ],
  "input": {
    "width": 128,
    "height": 128,
    "channels": 1,
    "normalization": "grayscale uint8 / 255",
    "resize": "bilinear"
  },
  "model_sha256": "3d393419b2e1606fb7f3eef32264413f1b83bd0b3d46bd8e0e01a59881a51ae9",
  "checkpoint_sha256": "447831c0c40c15d9ab491e34d4fd6b9a64bef603109fda179aba06722ee8bdd2",
  "plan_sha256": "3c14e1ce5cbc17c416c46f62fc86d7775353ee2f3f96253b38cb6b7f74db79f5",
  "counts": {
    "training": 368,
    "validation": 94,
    "heldout": 99,
    "heldout_video_groups": 47
  },
  "heldout": {
    "class_dice": [
      0.9756510885021429,
      0.6760492837824301,
      0.7156142060047825,
      0.5665771968724472,
      0.5226512367684967,
      0.490356281810876
    ],
    "foreground_macro_dice": 0.5942496410478064,
    "foreground_macro_dice_ci95": [
      0.5484885561940819,
      0.6244427592553611
    ],
    "ci_method": "1,000 source-video-group bootstrap resamples"
  },
  "always_background_baseline": {
    "class_dice": [
      0.9760630264235629,
      0.0,
      0.0,
      0.0,
      0.0,
      0.0
    ],
    "foreground_macro_dice": 0.0
  },
  "selected_validation_dice": 0.5785836865336746,
  "onnx_max_absolute_error": 0.0014500021934509277,
  "onnx_pixel_class_agreement": 1.0,
  "limitations": [
    "Trained and tested on one phantom collection, not a patient cohort.",
    "Patient performance, diagnostic accuracy, motion signs, and future outcomes have not been established.",
    "Source-video grouping does not establish independence of phantom acquisitions.",
    "Three duplicate images had discordant annotations; deterministic exclusions are recorded in the frozen plan.",
    "Pixel Dice and its video-group bootstrap interval cannot be interpreted as clinical sensitivity or specificity."
  ]
}